Biological Indicators – CSSD MCQ Questions and Study Notes
Biological indicators (BIs) are the most direct test of sterilizer effectiveness because they contain resistant spores. Practice free CSSD MCQ questions on this topic below.
Take the Biological Indicators practice test All 45 chapters
Learning objectives
- Understand the key concepts of this chapter
- Apply them in safe practice
Study notes
BIs use spores such as Geobacillus stearothermophilus for steam and hydrogen peroxide, and Bacillus atrophaeus for EtO and dry heat. Spores are chosen because they are the most resistant organisms.
A BI is placed in a challenge pack or process challenge device in the most difficult area of the load, processed, then incubated with an unprocessed control. Growth indicates failure. Steam sterilizers should be tested at least weekly and every load with implants.
Important points
- Test at least weekly, every implant load
- Always incubate a control
- Growth = failure
- Use the right spore
Key terms
Geobacillus stearothermophilus Bacillus atrophaeus PCD Control
Practice MCQs on Biological Indicators
Tap “Show answer” to check each answer and read the explanation. These practice questions are written for learning with AI assistance and may contain mistakes. Please report any error so it can be fixed.
Questions 1 to 10
1. BIs contain:
- Salt
- Enzymes only
- Dye
- Resistant bacterial spores
Show answer
Spores are hardest to kill.
2. Steam BIs use:
- E. coli
- Candida
- Staph aureus
- Geobacillus stearothermophilus
Show answer
Standard steam spore.
3. EtO BIs use:
- Geobacillus
- Yeast
- E. coli
- Bacillus atrophaeus
Show answer
Standard for EtO.
4. Positive BI result means:
- Dry load
- Success
- Sterilization failure
- Cooling
Show answer
Growth = failure.
5. A control BI is:
- Unprocessed and incubated
- Sterilized
- Discarded
- Frozen
Show answer
Confirms viability.
6. BI is placed:
- Anywhere
- In the most challenging load area
- In storage
- Outside the door
Show answer
Worst-case location.
7. Loads containing implants should be:
- Weekly only
- Not tested
- BI tested every load
- Monthly tested
Show answer
Implants need BI each load.
8. Minimum routine steam BI testing is:
- Monthly only
- Never
- At least weekly
- Yearly
Show answer
Per common standards.
9. When a BI fails:
- Recall/quarantine per policy
- Ignore
- Bury data
- Continue use
Show answer
Follow protocol.
10. BIs verify:
- Sterilizer lethality
- Packaging color
- Storage temp
- Cleaning
Show answer
Direct effectiveness check.
Questions 11 to 20
11. A biological indicator confirms:
- Item age
- Cleanliness
- Pack weight
- Killing of resistant spores
Show answer
India practice: the strongest test of sterilizer efficacy.
12. Steam BI vials are incubated at about:
- 20 °C
- 100 °C
- 56–60 °C
- 4 °C
Show answer
India practice: for Geobacillus.
13. A positive BI result means:
- Sterilization failure
- Success
- Cooling
- Dry load
Show answer
India practice: recall and investigate.
14. A control vial is:
- Discarded
- Unprocessed and incubated
- Frozen
- Processed only
Show answer
India practice: verifies spore viability.
15. Steam sterilizers are commonly BI tested:
- At least weekly (and every load with implants)
- Only when a failure is suspected
- Every load for non-implant items
- Once a year during the service visit
Show answer
India practice: follow NABH and IFU.
16. A negative BI result shows that:
- The tape failed
- The load is dirty
- The spores were killed in that test
- The printout is wrong
Show answer
Results are read per BI instructions.
17. In a steam load the BI or challenge device is often placed near the:
- Light
- Drain
- Door handle
- Ceiling
Show answer
The drain is a hard-to-reach spot.
18. Why must the BI incubator be at the right temperature?
- To keep staff warm
- Spores grow only under the correct conditions
- To change the color
- To dry the vials
Show answer
Wrong temperature gives false results.
19. How long should a BI be incubated?
- Until someone remembers
- As long as the BI maker states
- One year
- One minute
Show answer
Follow the instructions.
20. BI results should be:
- Thrown away
- Told verbally
- Hidden
- Recorded with date and lot details
Show answer
Records support traceability.
Related chapters
- Previous: Chemical Indicators
- Next: Bowie-Dick / Air Removal Testing
- Chapter 9: ETO Sterilization
- Chapter 10: Chemical Indicators
- Chapter 12: Bowie-Dick / Air Removal Testing
- Chapter 13: Sterilization Monitoring
Continue learning
CSSD interview questions Mock tests MCQ practice CSSD NABH checklist
Frequently asked questions
How many Biological Indicators MCQs are available?
This page has 20 free MCQs on Biological Indicators with answers and explanations. More are added over time.
Is Biological Indicators practice free?
Yes. You can practice without login or payment.
Are these official CSSD exam questions?
No. They are original practice questions written for learning.